Please use this identifier to cite or link to this item: http://repositorio.lnec.pt:8080/jspui/handle/123456789/1016928
Title: Analysis of flow rate measurement accuracy and traceability of flowmeters in field conditions using clamp-on ultrasonic flowmeters
Authors: Simões, C.
Ribeiro, A.
Almeida, M. C.
Covas, D.
Keywords: Flow measurement;Clamp-on ultrasonic flowmeter;Measurement uncertainty;Traceability;Measurement accuracy
Issue Date: Oct-2023
Publisher: IMEKO TC8
Abstract: This study examines the traceability and measurement uncertainty of in situ hydraulic calibration using clamp-on ultrasonic flowmeters as a reference. The procedure compares the equipment readings with the reference ones. Measurement uncertainty evaluation uses GUM formulation, considering the linearity conditions of the mathematical models applied. Experimental values are used to test the procedure and its suitability for actual cases where the expected accuracy needs to be achieved.
URI: https://repositorio.lnec.pt/jspui/handle/123456789/1016928
Appears in Collections:DHA/NES - Comunicações a congressos e artigos de revista

Files in This Item:
There are no files associated with this item.


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.